Description
Specifications
- Pressure: CPAP: 4 to 20 cm H2O (4 to 20 hPa) Resolution: 1 cm H2O (1 hPa) Dynamic Accuracy: ±5 cm H2O (±5 hPa)
- Pressure: EPAP: 4 to 20 cm H2O (4 to 30 hPa) Resolution: 1 cm H2O (1 hPa) Dynamic Accuracy: ±5 cm H2O (±5 hPa)
- Pressure: IPAP: 4 to 30 cm H2O (4 to 20 hPa) Resolution: 1 cm H2O (1 hPa) Dynamic Accuracy: ±5 cm H2O (±5 hPa)
- Rise Time : Settings 1 to 6 (where 1 = 0.1 sec and 6 = 0.6 sec) Resolution: 1 Accuracy: ±10% of setting (at R=50, C=20)
- Rate: 1 to 30 BPM Resolution: 1 BPM Accuracy: ±1BPM or ±10 % of setting whichever is greater over a 4 minute period. I-Time and Rate settings cannot allow I-Time to exceed expiratory time.
- I-Time: 0.5 to 3 sec Resolution: 0.1 sec Accuracy: ±0.1+10% of setting sec I-Time and Rate settings cannot allow I-Time to exceed expiratory time.
- Ramp Time: 0 to 45 min Resolution: 5 minAccuracy: ±10% of setting
- Ramp Start: 4 cm H2O to EPAP (4 hPa to EPAP) Resolution: 1 cm H2O (1 hPa)
- Patient Circuit Pressure: Range: 0 to 35 cm H2O (0 to 35 hPa) (continuous bar graph display)
- Rate: Range: 0 to 99 BPM
- Estimated Vt: Range: 0 to 4000 mL
- Estimated MV: Range: 0 to 99 L/min
- Patient Leak: Range: 0 to 200 L/min
- Apnea Rate: Range: 0 to 99 per hour
- High System Pressure: 5 cm H2O (5 hPa) over target IPAP pressure (Automatic)
- Low System Pressure5: 5 cm H2O (5 hPa) under target EPAP pressure (Automatic)
- Apnea: Range: 20, 40, 60 sec or Off
- Apnea Rate: Range: 5, 10, 20 apneas per hour or Off
- Nurse Call: Supports normally open or normally closed systems
- Dimensions: 5.5″ H x 11.4″ W x 14″ D (140 mm x 290 mm x 355 mm)
- Weight: 10 lb (4.5 kg)
- Power Input Rating: 100VAC / 230VAC and 50/60 Hz
- Battery Backup: 3.8 amp-hour (NiMH)
- Fuse: No replaceable fuse
- Nurse Call: Rated current: 0.280 A
- Rated voltage: 250 V