datascope accutorr plus

180,000.00

FEATURES

Light weight, portable with user-configurable technology

Easy-to-read large numbers

Dedicated controls which can be customized

Easily configurable:

–    Patient size

–    Interval time

–    Alarms

–    Inflation pressure

–    Audio level

–    Temperature format

 

Quick connect, single hose fittings

Unique flexible databases

IV pole/bed rail attachment

Lithium Ion Battery

Up to 8 hours run time without recharging

Over twice the life cycles of sealed lead acid batteries

Only 2 hours to fully recharge battery

Non-intrusive Blood Pressure (NiBP) Monitor

Stores up to 100 readings, instantly accessible

Optional – Predictive thermometer module

Optional – Printer module

Masimo pulse oximetry

Various flexible mounting options

SPECIFICATIONS

Systolic and Diastolic Pressure Readout

Number of Digits: 3

Accuracy*: Mean error < ±5 mmHg

Standard deviation < ±8 mmHg

Range: Adult Mode: 55 to 260 mmHg

Pediatric Mode: 55 to 160 mmHg

Neonatal Mode: 45 to 120 mmHg

 

Pulse Rate

Range: 35-245 BPM for Adult and Pediatric

70-245 BPM for Neonate

Display Resolution: 1 BPM

Accuracy: ±3 BPM or ±3%, whichever is greater

 

Inflation Source

This inflation source is capable of supplying sufficient air to bring a volume of 700 cc’s to a pressure of 300 mmHg in no more than 35 seconds. If the cuff is not inflated to the desired pressure within 60 seconds, then the cuff is vented and a retry cycle is initiated.

Leak Rate

With the bleed valve closed, the maximum pressure drop is 10 mmHg in 90 seconds measured with a 700 cc volume at a differential pressure of 250 mmHg.

Cuff Vent Rate

When the unit is vented, a volume of at least 700 cc’s is reduced from a pressure of 250 mmHg to a pressure of 20 mmHg in a maximum of 14 seconds.

Masimo SpO2

Range: 70-100% SpO2

Display Resolution: 1% SpO2

Display Update: < 4 seconds

SpO2 Accuracy Saturation during No Motion Conditions

Adults / Pediatrics: 70% to 100% ± 2 digits; <70% unspecified

Neonates (LNOP/LNCS): 70% to 100% ± 3 digits

SpO2 Accuracy Saturation during Motion Conditions

Adults / Pediatrics2: 70% to 100% ± 3 digits; <70% unspecified

Neonates3 (LNOP): 70% to 100% ± 4 digits

Neonates3 (LNCS): 70% to 100% ± 3 digits; <70% unspecified

SpO2 Response Time: 10 secs (Display Averaging Time is not user-selectable for Masimo. It will be set to the default value of 8 seconds.)

 CONTENTS

Qty 1 – Datascope Accutorr Plus Patient Monitor

Qty 1 – Lithium Ion Battery

Qty 1 – Blood Pressure Cuff

Qty 1 – Masimo Finger Probe

 

download brochure

Send an inquiry
Please, insert your name
Please, insert your email
Please, insert the message

Quantity: